Product Description
Established in the year 1994, we are becoming as one of the trusted manufacturers and suppliers of High Pressure Steam Sterilizer Horizontal. It is most commonly utilized in hospitals, laboratories, and clinical research centers. This sterilizer works on the principle of wet heat sterilization method to remove all the microbes & pathogens from the materials. It is completely suitable for sterilizing glassware, surgical equipments, media, and handmade swabs as well. High Pressure Steam Sterilizer Horizontal is generally operated at 121 degrees Celsius temperature and 15 pounds pressure for about 15 to 20 minutes.
Features:
-
It has the ability to kill resistant spores
-
It has steel alloy made jacketed structure
-
Equipped with digital screen and thermostat for temperature control
Size available 16x24
Size available 16x36
Size available 20x36
Size available 20x48
Advanced Control and Safety SystemsThis horizontal autoclave utilizes a PLC-based microprocessor controller, enhancing precision and automation in the sterilization process. The digital LED display offers real-time monitoring, while advanced safety features such as pressure interlock mechanisms, overheat protection, and a low water level cut-off ensure secure operation. These robust controls minimize operator error and support continuous, reliable performance across diverse applications.
Superior Build and Ergonomic DesignConstructed with either stainless steel 304 or 316, the chamber assures long-term corrosion resistance and durability. The mirror-polished interior and matte exterior finish facilitate easy cleaning and a professional appearance. Stainless steel perforated trolleys and spraying piping ensure efficient, rust-free operation. The floor-mounted unit, equipped with level adjusters, provides flexibility and stability for various installation environments in healthcare or research facilities.
FAQs of Horizontal High Pressure Steam Sterilizer:
Q: How does the PLC-based microprocessor controller enhance the sterilization process?
A: The PLC-based microprocessor controller ensures precise regulation of temperature, pressure, and time, automating sterilization cycles and minimizing manual intervention. This intelligent control system enhances reliability, repeatability, and traceability, critical for stringent hygiene standards in medical, pharmaceutical, and research settings.
Q: What materials are used for the chamber and other critical components?
A: The chamber is constructed from stainless steel 304 or 316 for optimal resistance to corrosion and thermal stress. Key components like the piping system, inner trolleys, and spraying system also use rust-free stainless steel, ensuring longevity and compliance with hygiene protocols.
Q: When should the sterilizers parameters be validated, and how is it done?
A: Sterilization parameters should be validated before initial use and periodically thereafter, especially after major maintenance or repairs. Validation is facilitated through provided test ports, allowing for independent measurement and verification using calibrated data loggers or chemical/biological indicators.
Q: Where can this sterilizer be installed, and what are its space requirements?
A: The unit is floor-mounted with level adjusters and requires enough space to accommodate its outer dimensions: 1600 x 700 x 900 mm, plus room for operator movement and trolley handling. It is suitable for hospitals, labs, pharmaceutical, and research institutes with access to a three-phase power supply and compressed air (6-7 bar).
Q: What is the typical process and working cycle for this steam sterilizer?
A: After loading the chamber with items on removable trolleys, users select the desired sterilization parameters on the digital controller. The system then delivers saturated steam (121C to 134C) for a preset period (0-99 minutes) with uniform steam distribution. Automatic safety valves and digital recording ensure monitored and secure operation throughout the cycle.
Q: How does the sterilizer ensure operator and equipment safety?
A: Safety is prioritized through automatic pressure interlocks, overheat protection, low water level cut-off, and seamless piping. These mechanisms prevent unsafe chamber opening, limit thermal risks, avoid dry-heating of elements, and minimize leakage or contamination, ensuring safe and efficient equipment usage.
Q: What are the benefits of using this horizontal high pressure steam sterilizer?
A: Benefits include fast, thorough sterilization of medical instruments and laboratory materials, effortless operation via digital controls, reduced downtime due to rapid temperature recovery, enhanced safety features, and durable construction. Its design and performance meet the rigorous demands of modern healthcare and research environments.